The Keithley 2510-AT is a 50 W bipolar thermoelectric cooler (TEC) controller that combines TEC drive and electrical measurement functions. It is designed for temperature control in continuous-wave laser-diode LIV (light-current-voltage) test systems and other tests of actively cooled optoelectronic devices. A compatible temperature sensor provides feedback to the digital proportional-integral-derivative (PID) control loop.

The 2510-AT adds automatic PID tuning to the base 2510 platform. Its autotuning function calculates the control-loop coefficients and offers minimum-overshoot and minimum-settling-time modes. It can regulate temperature, voltage, current, or thermistor resistance, with a temperature-setpoint range of −50 °C to +225 °C and TEC output of up to ±10 V DC and ±5 A DC.

Supported feedback devices include thermistors, RTDs, and compatible solid-state temperature sensors. Four-wire sensor-lead checking helps identify open or shorted connections. TEC current, voltage, power, and resistance measurements support evaluation of the thermal control system; the AC-resistance function uses pulsed current reversal and is available over the bus only. IEEE-488 and RS-232 interfaces support automated test integration.

Typical model applications include laser-diode modules, cooled photodetectors, thermal-optic switches, and temperature-controlled fixtures. Select the sensor, TEC load, cabling, and programmed limits for the intended device. The stability figures in the specifications below are manufacturer ratings under the stated sensor, ambient-temperature, and time-period conditions.

기술 정보

사양

TEC Output and Control

TEC output power
50 W bipolar output
TEC output range
±10 V DC at up to ±5 A DC
Control method
Digital PID control with independently programmable proportional, integral, and derivative gains
PID tuning
Automatic tuning; minimum-overshoot and minimum-settling-time modes
Control setpoints
Temperature, TEC voltage, TEC current, or thermistor resistance
Temperature-setpoint range
−50 °C to +225 °C
Temperature-setpoint resolution
0.001 °C (listed as ±0.001 °C in the manufacturer data sheet; this is resolution, not temperature accuracy)

Feedback and Measurement

RTD inputs
100 Ω and 1 kΩ
Thermistor inputs
100 Ω, 1 kΩ, 10 kΩ, and 100 kΩ
Solid-state sensor inputs
Compatible current-output or voltage-output temperature sensors
TEC measurement functions
Current, voltage, power, and resistance
AC-resistance function
Dual-pulsed current-reversal measurement; available over the bus only
Temperature-sensor lead detection
Four-wire open/short lead detection

Interfaces and Qualified Stability

Remote interfaces
IEEE-488 (SCPI 1995.0) and RS-232
Short-term setpoint stability
±0.005 °C rms over 24 hours, with a 10 kΩ thermistor. Manufacturer conditions: instrument and DUT in a regulated 25 °C ambient; Peltier and controller at 25 °C ±0.5 °C.
Long-term setpoint stability
±0.01 °C over 30 days, with a 10 kΩ thermistor. Manufacturer conditions: instrument and DUT in a regulated 25 °C ambient; Peltier and controller at 25 °C ±0.5 °C.
Mains input
90–260 V rms, 50–60 Hz; 75 W
Warm-up to rated accuracy
1 hour

판매 및 서비스

보증
6개월 보증
보정 상태
교정 증명서 또는 시험 보고서 제공 가능 (견적서에 명시된 바와 같이)
제조사에서 명시한 표준 액세서리
Input/output connector; User's manual.

왜 저희와 함께 일해야 할까요?

고객님과 함께 세부 사항을 확인합니다

사양부터 액세서리까지, 어떤 구성품이 포함되어 있는지 자세히 안내해 드립니다.

무엇을 구매하는지 정확히 파악하세요

저희는 해당 질환에 대해 알고 있는 정보와 확보된 기록을 모두 공유해 드립니다.

필요한 사항을 알려주세요

특별한 요구 사항이나 마감일이 있으신가요? 고객님께 중요한 사항에 집중해 드리겠습니다.

다시 시작할 필요는 없습니다

저희는 여러분의 질문과 이전 토론 내용을 기록해 두므로, 중단했던 부분부터 다시 이어가기 쉽습니다.

A closer look at 2510-AT

A closer look at 2510-AT
문의

이 제품에 대해 문의하기

요구 사항을 알려주시면 현재 제품 상황을 검토해 드리겠습니다.

이메일 주소나 전화번호를 입력하세요.

회사 이메일 주소를 입력한 다음, 해당 이메일로 전송된 6자리 코드를 입력해 주세요.